Describe event or problem: the medical center reported that all sizes of the deroyal neonatal feeding tubes (adaptor end) are stretching out once the extension set with clamp styas connected for a span of time.Due to the issue, the clinicians are having to use both hands to draw the residual back.Deroyal: the reported device has not been returned to deroyal at this time for evaluation.The investigation into the root cause is in process.It has been identified that after removing the extension set, if the connector is not given time to shrink back to its original size and the clinician uses a small syringe (1-10cc) then the connector may be loose and they will not be able to get a good draw.Testing is being performed to compare the use of 1-10cc syringes to the 20-60cc enteral syringes and extension sets.
|