Catalog Number 000000000000010220
Device Problem Insufficient Information (3190)
Patient Problems Bradycardia (1751); Death (1802); Dyspnea (1816); Sore Throat (2396); Low Oxygen Saturation (2477)
Event Date 05/30/2014
Event Type  Death  
Event Description
The customer reported that a patient in icu was undergoing a therapeutic plasma exchange(tpe) procedure in tandem with continuous veno-venous hemofiltration (cvvh).Approximately 15 minutes prior to completion of the tpe procedure, the patient was coughing,complained that his throat was hurting and felt short of breath.The patient's heart rate ando2 saturation decreased.The icu nurse called a code and attempted intubation.The patient then expired after the code was called.The disposable set was not available for return for evaluation because it was discarded by the customer.This report is being filed due to patient death, though at this time it is not suspected or alleged that the device was a contributory factor.
 
Manufacturer Narrative
Investigation: per the customer, the patient had been treated in (b)(6) at a sister facility wherehe was admitted and had 5 tpe procedures in tandem with cvvh.He had been intubated at thattime.He was septic with loss of consciousness changes, multiple system organ failure, respiratoryproblems and stroke.He was sent to the current facility for rehab.He was on o2 by cannula,breathing on his own, alert, and talking with his brother during the procedure.Investigation is in process.A follow-up report will be provided.
 
Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed.The signals in the run data file indicate that the spectra optia system operated as intended.Investigation is in-process.A follow-up report will be provided.
 
Manufacturer Narrative
Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.A service call was placed and a full machine checkout was performed.The machine is functioning per manufacturers specification.A review of the lot for similar reports was carried out and none have been reported.Multiple attempts to gather further information regarding the procedure and autopsy results remain unanswered.Root cause: this disposable set was unavailable for specific root cause analysis.The rdf and the service call show that the system operated as intended.A definitive root cause could not be determined.Based on the statements from the customer, possible causes include but are not limited to compromised patient physiology or related to how the cvvh procedure interacted with the apheresis procedure.Spectra optia apheresis system essentials guide offers the following procedural caution: "the spectra optia system has many safety features.However, a patient reaction can occur rapidly.Therefore, it is imperative that the operator monitor the patient and the system throughout the procedure.".
 
Manufacturer Narrative
Additional information: the medical director stated that the patient death was due to pulmonary embolus.
 
Manufacturer Narrative
Additional information: according to "therapeutic apheresis: a physician's handbook," the rate of adverse events during therapeutic apheresis is 4%-5%.Morbidity and mortality related to therapeutic procedures are greater in acutely ill patients treated in a hospital setting.A patient's death during an apheresis series is most often attributed to the underlying disease and is rarely directly related to the procedure.Root cause continued: the evidence from this investigation indicates that the device did not cause or contribute to the patient death.Literature review also indicates that a patient's death during anapheresis series is most often attributed to the underlying disease and is rarely directly related to the procedure.Based on the statements from the customer, the patient was septic and medically compromised with multi-system organ failure, respiratory problems, and stroke.
 
Manufacturer Narrative
This record was identified during a retrospective review of mdrs to identify records in which an event occurred, but the type of reportable event was not indicated appropriately in the mdr form.This supplement is being filed to modify information to align with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
jessica kim
10811 w. collins ave
lakewood, CO 80215
3032314812
MDR Report Key3900226
MDR Text Key4650237
Report Number1722028-2014-00252
Device Sequence Number1
Product Code LKN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK130065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/01/2016
Device Catalogue Number000000000000010220
Device Lot Number02W3225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
CVVH 05/30/2014
Patient Age00017 YR
Patient Weight177
-
-