Catalog Number 1011499-15
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem Foreign Body In Patient (2687)
Event Date 06/12/2014
Event Type  Injury  
Event Description
It was reported that during herculink elite stent deployment, the stent jumped forward, to an unintended site, within the renal artery.An armada dilatation catheter was able to fully appose the stent to the unintended area.Another herculink successfully covered the renal artery lesion.There was no clinically significant procedural delay reported.There was no additional information provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore, a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances.A query of the electronic complaint handling database revealed no other similar incidents reported from this lot.Based on the information reviewed, there is no indication of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Type of Device
RENAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key3900652
MDR Text Key4429011
Report Number2024168-2014-04182
Device Sequence Number1
Product Code NIN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 06/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Catalogue Number1011499-15
Device Lot Number3111261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-