|
Model Number 3389S-40 |
Device Problems
Bent (1059); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/09/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the lead was going to be used for an implant procedure, however upon visual examination by the clinician it was decided that the lead tip was too bent and the clinician did not want to use it for the procedure.The clinician was concerned this could cause inaccuracy during the placement procedure.The reporter also stated there was an extra bulge on the lead.Three days later it was reported that the bulge was at the distal tip.A different lead was used.
|
|
Manufacturer Narrative
|
(b)(4).Analysis of the lead, lot #va06yvt, found the distal end bent even though it was new out of the box.The distal end of the lead was bent at electrode 0.The lead had acceptable continuity and no shorts.
|
|
Search Alerts/Recalls
|
|
|