|
Model Number H749518080 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/11/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on (b)(4) 2014.During a percutaneous coronary intervention, an opticross was used on a mildly calcified lesion located at the left proximal anterior descending artery with 75 percent stenosis.Pre intravascular ultrasound (ivus) was performed before stenting to observe the vessel with this device, however "connect" and "disconnect" messages were displayed several times.When this device was used again during post ivus, "disconnect" message displayed repeatedly.Therefore, the device was exchanged to another opticross to complete the procedure.No patient complications reported and the patient's condition is good.However, device analysis revealed an open hole at the sheath lap joint section of the device.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluated by manufacturer: the complaint device was received for evaluation.Evaluation of the returned device found an open hole at the sheath lap joint section of the device.Fluid was leaking from the open hole at the sheath lap joint assembly when the catheter was flushed.The telescope assembly was able to properly pull back, advance, and retract.The ccp board and pins appeared normal and a good click sound was heard during insertion into the motor drive unit (mdu) system.No disconnect error message was observed during functional testing.During image characterization testing in the roller coaster model, a good square image appeared in the system and the product performed within specification.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Search Alerts/Recalls
|
|
|