(b)(4).Method: the complaint rt134 adult unheated breathing circuit was returned to fisher & paykel healthcare in (b)(4) for evaluation.The breathing circuit visually inspected, pressure tested, and immersed in a water bath to test for leaks.Results: the pressure test result for the complaint breathing circuit was out of specification due to an excessive leak.When the breathing circuit was immersed in the water bath, the leak was observed at the connection between the lid and the bowl of the water trap.A lot check could not be performed as no lot information was provided.Conclusion: the breathing circuit water trap consists of a bowl and a lid, which can be separated to allow the caregiver to empty the water trap.All breathing circuits are pressure tested for leaks during production and those that fail are rejected.This suggests any leak must have developed after the rt134 adult unheated breathing circuit was released for distribution, during transport, storage or use, possibly by distortion of the water trap when the bowl was connected.Our user instructions that accompany the rt134 adult unheated breathing circuit state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate alarms." hospital staff correctly checked the breathing circuit before patient use, which is in line with our user instructions.
|