• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problem Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/13/2014
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice automated pd set with cassette, the device was found to be cut along the edge of the cassette sheeting weld located on the lower right corner of the cassette.No additional information is available.This is report 5 of 8.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned to baxter and the evaluation is complete.A visual inspection was performed with cuts in the cassette sheeting noted.A leak test was performed with a leak from the cassette noted.The device also failed to meet pressure testing requirements.Clamp function testing and clear passage testing were performed with no issues noted.The sample was not run on the homechoice machine due to the physical failure found.The cause of the condition was not determined.A capa was opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3903594
MDR Text Key19925279
Report Number1416980-2014-20974
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2018
Device Catalogue NumberL5C4531
Device Lot NumberH13L22013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2014
Date Manufacturer Received05/13/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-