• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q; OPHTHALMIC EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Model Number 8065990601
Device Problem Energy Spectrum Incorrect (1210)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/28/2014
Event Type  malfunction  
Event Description
A technician reported a case where after completing one percent of the lasik treatment, secondary energy too high code displayed, and indicated the procedure was completed.Reporter indicated they contacted the field service engineer who instructed them to restart the system and perform multiple energy checks, which cleared the code and they had no issues.Reporter relayed the doctor decided to continue and finished the surgery day.Technician spoke to doctor and treatment outcome was as good.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable info becomes available.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 9105 8
GM   91058
Manufacturer Contact
janet moran
6201 south freeway, r3-48
fort worth, TX 76134
8176152742
MDR Report Key3904772
MDR Text Key16304912
Report Number3003288808-2014-00913
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8065990601
Device Catalogue Number8065990601
Other Device ID NumberV4-1.02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-