(b)(4).Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions which verified pacing, sensing, impedance measurements and recording functions; the device passed electrical testing.Notes related to left ventricular (lv) lead programming were found in the device memory and when longevity calculations were performed based on those measurements the device met longevity.The field observations were unable to be confirmed by analysis.
|