Model Number H749518080 |
Device Problem
Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/17/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on analysis completed on (b)(4) 2014.It was reported that the transducer did not move back to its initial position.The opticross¿ imaging catheter was used to visualize an unspecified lesion.It was noted during percutaneous coronary intervention, post intravascular ultrasound, the transducer did not move back to its initial position.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition is good.However, returned device analysis revealed an open hole at the sheath lap joint assembly.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: the complaint device was returned for evaluation.The telescope assembly was not able to properly pull back, advance, or retract.An open hole was observed at the sheath lap joint section of the device.Fluid was leaking from the open hole at the sheath lap joint assembly when the catheter was flushed.During image characterization testing, no image appeared in the system due to electrical open at proximal.No imaging core windup was found within the telescope section of the device.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
|
|