It was reported that during a da vinci surgical procedure, a derailed cable was observed on the permanent cautery hook instrument.There was no report of fragments falling into the patient.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.
|
The instrument was returned and evaluated.Failure analysis investigation was unable to confirm the reported complaint of a derailed cable.The instrument does not have cables or wires outside of the housing; thus, no cables or wires are visible.Failure analysis investigation found that the instrument's main tube was broken at the connection with the clevis.There appeared to be material missing from the main tube as a result of the damage.It was concluded that the damage was likely due to mishandling/misuse.No other damage was found.The endowrist® instruments instructions for use (ifu) specifically states: general precautions and warnings handle instruments with care.Avoid mechanical shock or stress that can cause damage to the instruments.The customer reported complaint does not itself constitute a mdr reportable event; however, the damage to the main tube, found during failure analysis investigation, could likely cause or contribute to an adverse event, if the malfunctions were to recur.
|