It was reported that the vns patient underwent surgery on (b)(6) 2014 to explant his generator and lead.The patient¿s device was explanted because the patient became seizure-free after having resective brain surgery.The explanted generator and lead were returned to the manufacturer for analysis.During visual analysis of the returned lead portion, abraded openings were found on the outer and inner silicone tubes which most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids found inside the outer and inner silicone tubes.With the exception of the abraded tubing openings, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.The setscrew marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Note that since the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Analysis of the returned generator found that the elective replacement indicator was set to ¿yes¿.An open can measurement of the battery voltage confirmed that the eri flag had been properly set; the battery was partially depleted.Based on the bench analysis and the electrical test results, the device exhibited current consumption rates that were within specification; thereby, demonstrating normal battery depletion.With the exception of the eri parameter (eri flag was set to ¿yes¿ due to a low battery condition), the device performed according to functional specifications.Therefore, the electrical performance of the generator, as measured in the pa lab, will be used to conclude that no abnormal performance or any other type of adverse condition was found.
|