Product condition received: the product was returned with the membrane completely unfolded and blood on the exterior of the catheter.No blood was observed inside the balloon, catheter, or extracorporeal tubing.One kink was found near the y-fitting approx 76.2cm from the tip.Product evaluation: an underwater leak test of the balloon, catheter, y-fitting, and extracorporeal tubing was performed and no leaks were detected.The iab was placed on the cs300 pump and pumped for two hours which represents one complete autofill cycle.The kink was induced and the iab pumped normally and no alarm sounded from the pump.Conclusion: the returned product was tested and reported alarms could not be duplicated.(b)(4).
|