The cpb circuit included a fx15 oxygenator and a sarns centrifugal pump head.There were no functional or behavioral issues observed during set-up and priming of the cpb circuit.On the initiation of cpb: blood flow, pump speeds, and oxygenator function were within normal limits and there were no signs of functional issues.The patient was slowly cooled and the aorta was cross clamped ad the 1st dose of cardioplegia was administered.The blood flow was about 5 1/min when the cross-clamp was applied.Over the next 5-10 minutes the blood flow gradually dropped in spite of raising pump speeds.Eventually, only 2.5 1/min of blood flow could be provided at full pump speed (3500 rpm).The circuit line pressure after the oxygenator was 148 mmhg.As the blood flows were dropping gradually the pao2 was also dropping to less than 100 mmhg.The cv surgeon was alerted to the issues and the team elected to change out both the oxygenator and the sarns centrifugal pump.Before the change out was performed, the patient was further cooled to 30 degrees c.When the temperature reached 30 degrees c, the arterial and venous lines were clamped and cpb was halted.The components were removed from the cpb circuit and a like oxygenator and centrifugal pump head were inserted, primed, and debubbled.Cpb was re-initiated and the procedure was completed as scheduled.The time off cpb was 4 minutes and the estimated blood loss was 200 ml.The estimated delay in the procedure was 5 minutes.The patient was weaned from cpb without issue and there was no harm observed in the operating room and in the 1st 12 hours post-op.As the patient was waking up in the icu, the patient was alert and responding to commands, and moving all limbs.About 1:00 am on (b)(6), 2014 the patient began to exhibit stroke-like symptoms and these included left sided weakness and slurred speech.The pump record was reviewed.Between 12:15 -12:45 pm.The blood flow dropped from 5.0 1/min to 2.5 1/min in spite of increasing the pump motor rpm tp 3500 rpm.During this same period, the pao2 dropped from 205 mmhg to 85 mmhg, and the svo2 dropped from 84% to 70% in spite of dropping the temperature to 30 degrees c.There is no indication of a structural issues with the unit and no indication of functional issues with the centrifugal pump head as flow was being produced and no signs/symptoms of decoupling or other functional failures.The pump record did detail that the original fx 15 unit was replaced with a fx 25 sized unit.
|