Catalog Number R5C4479C |
Device Problems
Component Falling (1105); Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/19/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that a homechoice device experienced a system error 2240 alarm (air in line/set).This event occurred during dwell cycle one of three of peritoneal dialysis therapy while the patient was connected.During troubleshooting, the patient stated that the supply bag fell off because of a loose connection.The technical service representative advised the patient to make sure that the clamp is closed on the bag when connecting and advised the patient to restart therapy with new supplies.There was no patient injury or medical intervention reported.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).An alarm indicative of a potential malfunction of the disposable cassette was identified.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).As the device was not returned and the lot number was unknown, a device analysis could not be completed.The cause of the reported alarm was due to a loose connection between a cassette supply line and a supply bag.The cause of the loose connection could not be determined.Should additional relevant information become available, a follow-up report will be submitted.
|
|