During a cardiac arrest of a trauma patient, patient was receiving prbc via a 3m ranger pressure infuser model 145, 3m ranger blood warming system model #245, using 3m tubing.During the infusion the blood tubing ruptured causing blood to spew all over the patient's room and 3 staff members assisting with the resuscitation.The post event review revealed no operator error.Patient had received 2 units prbc's & 1 unit normal ns in the ed prior to transfer to icu from a different ranger pressure infusion system without incident.Tubing for both of these events was from the same lot# (hx7355) and expiration date (b)(6) 2016.Following this malfunction, tubing from a different lot number was successfully used with the same ranger unit to infuse an additional unit of prbc's.
|