The device referenced in this report has not been returned to olympus for evaluation.The exact cause of the user's experience could not be conclusively determined at this time.However, based on the information provided, the reported phenomenon appears to be due to user error, as the device was used off-label.The bml-v442qr instruction manual intended use section states: "this instrument has been designed to be used with an olympus endoscope for crushing calculi inside the bile duct.Do not use this instrument for any purpose other than its intended use".If additional information or if the device is received at a later time this report will be supplemented.
|