MAUDE Adverse Event Report: BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL) CARTO® 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL) CARTO® 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number M-4800-01
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problems Low Blood Pressure/ Hypotension (1914); Perforation (2001); Cardiac Tamponade (2226)
Event Date 06/24/2014
Event Type  Injury  
Event Description
It was reported that a patient, (b)(6), female, underwent an idiopathic ventricular tachycardia (vt) ablation procedure with a carto 3 mapping system and suffered a cardiac tamponade which required pericardiocentesis.The patient had no reported medical history that contributed to the event.During the procedure, the carto 3 mapping system froze as the physician attempted to map the location of the catheters inside of the heart.To troubleshoot the software issue, the workstation was rebooted; the case was reopened and continued.There was a noticeable map shift error of at least 10 mm that occurred when attempting to create maps utilizing cartosound and paso software on the carto mapping system.The physician then noticed that the patient¿s blood pressure was decreasing and diagnosed a pericardial effusion using intracardiac echocardiography (ice) and believes that one of the catheters in place must have caused a perforation.The procedure was aborted.A pericardiocentesis was performed to remove 150 ml of fluid and a pigtail catheter was put in place to allow drainage.The patient did not require hospitalization and fully recovered with no residual effects.The physician¿s opinion regarding the cause of this adverse event is that this event may be due to a diagnostic medtronic marinr quad catheter that was also being used during the event, as he believes it may have perforated the heart.Nonetheless, this event is being reported under the carto 3 system because such reported map shift may contribute to potential patient risk.This event was also reported to the fda under the bwi thermocool catheter (as a separate report).The bwi failure analysis has begun but is not completed at this time.
 
Manufacturer Narrative
When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Concomitant products: stockert 70 (serial # (b)(4), model# m-5463-01).Coolflow pump (serial # (b)(4), model# m-5491-02).Soundstar® 3d diagnostic ultrasound catheter (model # m-5723-12).Decanav¿ electrophysiology catheter (model#, d-1283-02-s, lot number unk).Thermocool® smarttouch® bi-directional navigation catheter (model # d-1327-04-s, lot # 16086430m) (b)(4).
 
Manufacturer Narrative
(b)(4).It was reported that a patient, (b)(6) female, underwent an idiopathic ventricular tachycardia ablation procedure with a carto 3 mapping system and suffered a cardiac tamponade which required pericardiocentesis.The investigational analysis has been completed.Carto 3 system was replaced and suspicious system was sent to htc for investigation: the customer complaint was not confirmed.The system passed atp at service jig utility station and was tested at aquarium station with a few types of catheters (uls and st).Workstation was tested by sw department and csd to dig up some more information from this case and it was found that the data has been deleted on prior to shipment at htc to fulfil the hipaa requirement.No dump file or recordings were saved on the workstation, so further investigation couldn't be performed.The history of customer complaints associated with carto 3 system # (b)(4) was reviewed.Out of 4 additional reported complaints there was no any additional complaint that may be related to the reported issue.A dhr review was performed by the manufacturer and no anomalies were noted in manufacturing or servicing of this equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO® 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL)
4 hatnufah st.
yokneam 20692
IS  20692
Manufacturer (Section G)
BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL)
4 hatnufah st.
yokneam 2069 2
IS   20692
Manufacturer Contact
jaime chavez
15715 arrow highway
irwindale, CA 91706
9098398483
MDR Report Key3935106
MDR Text Key4599250
Report Number3008203003-2014-00049
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberM-4800-01
Device Catalogue NumberFG540000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/24/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other; Required Intervention;
Patient Age48 YR
-
-