• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. PRESSURE INFUSION BAG; INFUSOR, PRESSURE, FOR I.V. BAGS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. PRESSURE INFUSION BAG; INFUSOR, PRESSURE, FOR I.V. BAGS Back to Search Results
Catalog Number PIB500
Device Problems Air Leak (1008); Loss of or Failure to Bond (1068)
Patient Problem Misdiagnosis (2159)
Event Date 05/15/2014
Event Type  Injury  
Event Description
It was reported that a patient's blood pressure appeared to drop slightly on the monitoring equipment.The patient received a low dose of medication for drop in pressure.It was identified that the pressure infusion was part of the pressure monitoring system.The measured drop in blood pressure was attributed to the loss of pressure in the bag.Multiple attempts were made to get clinical details from a medical professional without success.No further harm or injury to the patient was reported.
 
Manufacturer Narrative
The user reported a second lot number (i677857) with a manufacturing date of april 18, 2014 and an expiration date of march 31, 2014.Device evaluation: the suspect device has not been returned for evaluation.A follow up report will be submitted when the evaluation has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE INFUSION BAG
Type of Device
INFUSOR, PRESSURE, FOR I.V. BAGS
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
south jordan UT
Manufacturer Contact
jerry mcphie
1600 west merit pkwy.
south jordan, UT 84095
8012084491
MDR Report Key3940692
MDR Text Key4530624
Report Number1721504-2014-00118
Device Sequence Number1
Product Code KZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Catalogue NumberPIB500
Device Lot NumberI668545
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-