• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problems Component Falling (1105); Disconnection (1171); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2014
Event Type  malfunction  
Event Description
It was reported that a system error 2240 (air in set) alarm occurred on a home choice (hc) device during dwell three of four.The patient was connected to the homechoice at the time of the alarm.The technical service representative explained the alarm and assisted the patient in clearing the alarm.During the troubleshooting, it was determined that the solution bag fell from an unstable surface and disconnected.The patient was to start therapy over with new supplies.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available at this time.
 
Manufacturer Narrative
(b)(4).The device was not returned and the lot number was unknown; therefore, a device analysis could not be completed.However, a bag falling/disconnecting is a known cause of this alarm.Proper user instructions are addressed in ¿the homechoice and homechoice pro apd systems patient at-home guide¿ which is shipped with every homechoice device.The guide instructs the user to place the solution bags on a flat, stable surface and it warns that falling bags can result in a disconnection or leak.The guide also instructs the user to check all disposable set connections for a secure fit before beginning therapy and provides step-by-step instructions for connecting the solution bags.A formal review of the label for the product family will be conducted.If additional relevant information is obtained, then a follow-up mdr will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3946653
MDR Text Key4536922
Report Number1416980-2014-23444
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Was Device Available for Evaluation? No
Date Manufacturer Received06/28/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age84 YR
-
-