• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO TPS HANDPIECE CORD; DRIVER, WIRE, AND BONE DRILL, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO TPS HANDPIECE CORD; DRIVER, WIRE, AND BONE DRILL, MANUAL Back to Search Results
Catalog Number 5100004000
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 07/03/2014
Event Type  malfunction  
Event Description
It was reported that during testing conducted at the manufacturer facility the tps handpiece cord caused a bias current message to be displayed on the console, signaling a condition occurred in which the device has the potential to run without user activation.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
During the device evaluation, the technician determined the cable¿s internal wiring was damaged.The device will be discarded by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPS HANDPIECE CORD
Type of Device
DRIVER, WIRE, AND BONE DRILL, MANUAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3950096
MDR Text Key4646129
Report Number0001811755-2014-02611
Device Sequence Number1
Product Code DZJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K943540
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5100004000
Device Lot Number12041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2014
Date Manufacturer Received07/03/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-