Final device investigation found that the device was returned in good visual condition with a clip in the jaw.The jaw appeared to be in alignment, and the clip retainer, push fork and jaw clearance were acceptable.Upon eval, the 13 remaining clips were fired.The first 4 clips loaded properly and had proper pinch and alignment.After th 4th clip, the device failed to load consistently, taking 2-3 actuations of the handle before a clip would load and fire.When fired, the clips had proper pinch and alignment.After the last clip was fired, the device locking mechanism did not engage.Further examination determined that there was a broken spring which caused the loading problem and failure to lock.It was not known whether the spring broke during device eval.The device history record was reviewed, and no discrepancies were noted.
|