Catalog Number 530.610 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/07/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during an orthopedic surgery, it was discovered that the battery oscillator device was losing power.The reporter claimed that all batteries were tested and worked fine.There were no delays to the planned surgical procedure as a spare identical device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that one of the battery contacts was damaged.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to worn components due to usage wear over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|