Catalog Number 511.300
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported when using the radiolucent drive under x-ray light, the operational area was not clearly visible.Customer reported "view as if metal was visible in the body".It is unknown if there was patient involvement during this event.This is report 1 of 1 for com-(b)(4).
 
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.(b)(6).The 510k#: device is not distributed in the united states, but is similar to device marketed in the usa.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional narrative: service history review of the device has been performed.The review indicates that the device has not been serviced during the past 6 months.There is no information relevant to the current complaint issue.The investigation is based on the service & repair record (srr) received.The device failure reported by the customer could be confirmed.The service history review shows no previous service conditions relevant to the current complaint issue.The service technician noted the following action taken: repair.The device has been serviced and functional testing has been performed in accordance with the service & repair procedure se_037288.During the pre-repair diagnostic assessment the service technician identified the following failure: coupling tool side worn out.The probable root cause of the event was determined to be normal wear.The device was repaired and returned to the customer if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIOLUCENT-DRIVE MARK II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3953723
MDR Text Key4611958
Report Number3009450871-2014-10275
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/09/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-