Catalog Number 530.100 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unspecified surgical procedure, it was discovered that the power drive device stopped working.There were no delays to the surgical procedure as a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of the event was unknown; however, the reporter stated that the event was discovered on (b)(6) 2014.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device did not run when power was applied.Therefore, the reported condition was confirmed.The assignable root cause was determined to be most likely due to motor assembly or electronic failure due to normal wear and usage over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|