• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES OBERDORF POWER DRIVE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF POWER DRIVE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.100
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2014
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure, it was discovered that the power drive device stopped working.There were no delays to the surgical procedure as a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of the event was unknown; however, the reporter stated that the event was discovered on (b)(6) 2014.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device did not run when power was applied.Therefore, the reported condition was confirmed.The assignable root cause was determined to be most likely due to motor assembly or electronic failure due to normal wear and usage over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER DRIVE
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3955358
MDR Text Key17157405
Report Number8030965-2014-10417
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.100
Device Lot Number34156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received09/10/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-