Catalog Number 178833
Device Problems Fluid/Blood Leak (1250); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/26/2014
Event Type  malfunction  
Event Description
The customer reported the instrument was generating partial aspiration errors and approximately 50 mls of bloody fluid leaked below the needles assembly inside the coulter lh 500 hematology analyzer.The leak was not contained within the instrument.There was no exposure to open wounds or mucous membranes.The customer was wearing personal protective equipment (ppe) consisting of gloves and a lab coat at the time of the occurrence.No erroneous results were generated in connection with the reported event.There was no death, injury, or affect to user or to patient treatment attributed to this event.
 
Manufacturer Narrative
A beckman coulter (bec) field service engineer (fse) was dispatched to assess the instrument.The fse observed a leak through vacuum tubing connected to the needle bellows.The fse replaced the tubing and the instrument performed without leaks or errors.The fse replaced an actuator in the needle drain manifold as incidental service.The repairs were verified and the instrument was returned to normal operation.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 500 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key3956320
MDR Text Key20017256
Report Number1061932-2014-01704
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042724
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number178833
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-