Upon receipt at medtronic's quality assurance laboratory, visual analysis confirmed the disconnection.There appeared to be damage to the tubing and sump at the interface where the disconnection occurred.After cleaning, further visual examination under magnification confirmed the presence of adhesive residue on the connection.The product has been sent to the contract manufacturer of this device for further examination.Upon completion of the investigation by medtronic's quality engineers and the contract manufacturer, a supplemental report will be filed.(b)(4).
|