A steris service technician arrived onsite and reviewed the cycle printout subject of the reported event.The cycle printout confirmed the cycle completed successfully with no alarms or aborts.In addition, the indicator present in the load evidenced passing results.The technician noted the cycle printout included a message which read "load test repeated".This message suggests that moisture is present in the load.The technician inspected the unit and found it to be operating properly; no issues were noted.The employee was not wearing proper ppe during the time of the reported event, specifically gloves.The operator manual states (99.1-3), "failure to thoroughly clean, rinse and dry articles to be sterilized could result in an ineffective sterilization cycle." the operator manual states (pp.6-29), "unload sterilization unit- steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical-resistant gloves when using the sterilization unit." while onsite the technician discussed the importance of wearing proper ppe and the proper use and operation of the sterilizer.
|