On (b)(6) 2014, dentsply was contacted by an attorney from the u.S.Attorney's office, to obtain sterilization instructions for midwest low speed handpieces.The attorney stated the (b)(6) uses gas to sterilize their low speed handpieces and asked if it was acceptable.Dentsply explained that we have validated steam sterilization and cannot verify if gas sterilization is acceptable.The attorney explained that a patient of the (b)(6) hospital filed a claim alleging that he contracted (b)(6) following a dental treatment where a midwest low speed handpiece and disposable prophy angle where used.The attorney further stated that the (b)(6) does not believe that the patient contracted (b)(6) from the dental treatment, rather, they believe that the patient contracted the virus as a result of his known drug addiction.
|
While it may be possible the patient contracted (b)(6) as a result of his known drug addiction, it is impossible to ascertain as the device was not sterilized according to the directions for use; therefore, because it could be possible that the device contributed to a serious illness in this event, it is reportable per 21 cfr part 803.The device was not returned for evaluation and the serial number was provided for a dhr review.
|