• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO PERFUSION SYSTEM 8000; 8K (PUMP, ROLLER TYPE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO PERFUSION SYSTEM 8000; 8K (PUMP, ROLLER TYPE) Back to Search Results
Model Number 16402
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the occlusion on the roller pump was hard to turn.As a result, an alternate device was employed.The surgical procedure ws completed successfully.There were no delays, no blood loss, or no adverse consequences to the pt.
 
Manufacturer Narrative
This complaint was found during the same case set-up as mdr #1828100-2014-00508.The field service rep (fsr) will repair the unit during the (b)(4) preventative maintenance (pm) schedule.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO PERFUSION SYSTEM 8000
Type of Device
8K (PUMP, ROLLER TYPE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key3966077
MDR Text Key15104725
Report Number1828100-2014-00512
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-