MAUDE Adverse Event Report: LIFE TECHNOLOGIES HOLDINGS PTE. LTD. APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR; INSTRUMENTATION FOR CLINICAL MULTIPL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE TECHNOLOGIES HOLDINGS PTE. LTD. APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR; INSTRUMENTATION FOR CLINICAL MULTIPL Back to Search Results
Catalog Number 4407205
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/24/2013
Event Type  malfunction  
Event Description
A customer reported noisy amplification plot and component plot on the instrument applied biosystem 7500 fast dx (cat no.4407205) with serial no.(b)(4), resulting in incorrect, wavy results.No pt involvement reported.Company event tracking no.(b)(4).
 
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.Repair completed by changing the lamp and performing optical alignment and re-optical calibration.The repaired instrument passed all specifications.This device is used to obtain results for a diagnosis.Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1.4 is an real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes.The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR
Type of Device
INSTRUMENTATION FOR CLINICAL MULTIPL
Manufacturer (Section D)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
singapore
SN 
Manufacturer (Section G)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer Contact
kelli tanzella
3175 staley road
estate road 3, # 05-03
grand island, NY 14072
7167743122
MDR Report Key3978856
MDR Text Key4822221
Report Number3003673482-2014-00027
Device Sequence Number1
Product Code NSU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/24/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4407205
Is the Reporter a Health Professional? No
Date Manufacturer Received04/24/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-