• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE TECHNOLOGIES HOLDINGS PTE LTD APPLIED BIOSYSTEMS 7500 FAST DX REAK TIME PCR; INSRUMENTATION FOR CLINICAL MULTIPL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE TECHNOLOGIES HOLDINGS PTE LTD APPLIED BIOSYSTEMS 7500 FAST DX REAK TIME PCR; INSRUMENTATION FOR CLINICAL MULTIPL Back to Search Results
Catalog Number 4406985
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2014
Event Type  malfunction  
Event Description
A customer reported a heater block issue on the instrument applied biosystems 7500 fast dx (catalog no.4406985) with serial no.(b)(4), resulting in incorrect results from an assay run.No pt involvement reported.(b)(4).
 
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.The repair was completed by replacing the sample block.The instrument passed all specification tests after the sample block was replaced.This device is used to obtain results for a diagnosis.Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1.4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes.The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures and on use of the analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLIED BIOSYSTEMS 7500 FAST DX REAK TIME PCR
Type of Device
INSRUMENTATION FOR CLINICAL MULTIPL
Manufacturer (Section D)
LIFE TECHNOLOGIES HOLDINGS PTE LTD
singapore
SN 
Manufacturer (Section G)
LIFE TECHNOLOGIES HOLDINGS PTE LTD
Manufacturer Contact
kelli tanzella
3175 staley rd
grand island, NY 14072
7167743122
MDR Report Key3978887
MDR Text Key18064203
Report Number3003673482-2014-00020
Device Sequence Number1
Product Code NSU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4406985
Is the Reporter a Health Professional? No
Date Manufacturer Received01/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-