A visual examination confirmed the reported event.The tip of the instrument is sheared off; the remaining portion of the tip is slightly deformed in a clockwise torquing/twisting manner.Due to the condition of the returned plug driver a functional evaluation was not performed.The rep stated this was a removal case and therefore deformation should be seen in a counter-clockwise direction; this instrument tip may have previously been damaged/deformed during tightening of the plug.A review of the device history records indicated that there were no dimensional, functional, or material anomalies reported at inspection.The probable underlying root cause for the damage is excessive torque forces leading to loss of mechanical integrity and ultimately instrument deformation and fracture during usage of the device.
|