(b)(4).The analysis results found that the er420 instrument was returned in good visual condition with a clip jammed between the top shroud and the jaws.The clip was removed in order to perform functional testing.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing the device for functionality the device was cycled, fed, and formed seven clips as intended and it ejected ten clips.Finally, it locked out as intended.The batch record was reviewed and no anomalies were noted during the manufacturing process.
|