CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/15/2014
Event Type  malfunction  
Event Description
It was reported that the connection between the usb adaptor and usb port for the tablet device was loose.The tablet device and usb adaptor have not been returned to date.
 
Event Description
The programming tablet was received for analysis.Analysis of the tablet was completed on 08/12/2014.The cause for the serial cable anomaly is associated with a broken wire connection in the serial cable db9 hood assembly.Once the wire was soldered on to the pcb, no further anomalies associated with the tablet performance were noted during testing using the ac adapter or the main battery with a full charge.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3987729
MDR Text Key4755134
Report Number1644487-2014-01966
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 10.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-