• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IPG MFG SWITZERLAND SAPHIR 3; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IPG MFG SWITZERLAND SAPHIR 3; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE Back to Search Results
Model Number SP640U
Device Problems Failure to Interrogate (1332); Loss of Data (2903); Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2014
Event Type  Injury  
Event Description
It was reported that the device could not be interrogated, and that the programmer indicated that data was corrupt.Electrical reset was also noted.The device was explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Evaluation summary: analysis of the device revealed normal battery depletion.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.This model number is not approved for distribution in the united states, however, it is similar to a device marketed in the u.S.The event is being reported due to an alleged malfunction.(b)(4).
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAPHIR 3
Type of Device
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3989853
MDR Text Key4885030
Report Number9614453-2014-01840
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P990001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/28/2007
Device Model NumberSP640U
Device Catalogue NumberSP640U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00086 YR
-
-