Catalog Number EMC9663C |
Device Problems
Component Missing (2306); Chemical Spillage (2894)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was destroyed by the customer.A review of all batch record documents was performed with no issues noted during the manufacturing process.There were no deviations from standard procedure and no exceptions related to the reported condition were noted.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink chemotherapy set was missing a ¿lead¿.This was noted when a nurse was priming the set with sodium chloride, and led to the sodium chloride spilling on the nurse.There was no report of medical intervention or adverse event associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Complaint no: (b)(4).The device was not received for evaluation; therefore, a device analysis could not be performed.A batch review will be performed.Should additional relevant information become available, a supplemental report will be submitted.
|
|