Model Number MODEL 100
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Date 07/22/2014
Event Type  malfunction  
Event Description
Complainant alleged that during a training session, the middle portion of the autopulse® platform display was dark and then unreadable signs were seen.There was no report of any patient involvement.No further details were provided.
 
Manufacturer Narrative
The autopulse platform (s/n (b)(4)) was returned to the manufacturer for analysis.Visual inspection of the returned platform shows no physical damage to the platform.A review of the autopulse archive was performed and the archive data shows that user advisories (uas) 45 (not at "home" position after power-on/restart) and 12 (lifeband® not present) occurred on the reported event date of (b)(6) 2014.However, these user advisories are not related to the reported complaint.Functional testing was performed and the reported issue could not be reproduced.The autopulse platform ran for 5 minutes with the test mannequin and additional minutes with the large resuscitation test fixture (lrtf) and no problems were observed.The platform passed functional testing.Based on the investigation, the lcd display will be replaced as a precaution.In summary, the reported complaint was not confirmed during visual inspection or during functional testing.The user advisories noted in the archive review are unrelated to the reported complaint.The root cause for ua45 and ua12 could not be determined.Per the autopulse® resuscitation system model 100 user guide (pn: 12555-001), if the driveshaft is not at its home position when the autopulse is powered on, a user advisory (45) will occur.This user advisory will persist until the driveshaft is returned to its home position.User advisory 12 is an indication that the autopulse has detected that the lifeband is not properly installed.Upon replacement of the lcd display, the platform was re-evaluated through functional testing and the platform passed all testing criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION SYSTEM MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION, INC
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key4004199
MDR Text Key4773352
Report Number3010617000-2014-00410
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/22/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-