(b)(4).Method: the complaint mr290 chamber was returned to fisher & paykel healthcare in (b)(6) and was visually inspected.Results: visual inspection revealed a break in the feedset tubing approximately 480 mm from the spike connection.The feedset tubing was observed as being compressed at the point of the cut, and the edges of the cut have been sealed.A lot check revealed no other complaints of this nature for lot 131022.Conclusion: we were unable to definitively determine the cause of the feedset damage, however, due to the location of the cut, it would not have been possible for the damage to have occurred whilst the feedset was still wound on the feedset winder.We have conducted extensive testing of the mr290 chamber, with particular emphasis on feedset breaks.The specification for the chamber requires that the feedset tube should have a breaking strain of 30 newtons.During production, pull testing of the feedset strength at both spike and dome end is performed every hour on mr290 chambers from each production line.If any faults are detected the whole batch is placed on hold for investigation.Additionally all chambers are pressure tested before they leave the production line and any holes or leaks in the feedset are identified during this process.Chambers that fail any of these tests are discarded.This indicates that the feedset became damaged after distribution.The user instructions which accompany the mr290 autofeed humidification chambers state the following: - "set appropriate ventilator alarms." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." (b)(4).
|