U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problems Device Displays Incorrect Message (2591); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/16/2014
Event Type  malfunction  
Event Description
The customer reported r (review result) flagging occurred on a patient sample when using the unicel dxh 800 coulter cellular analysis system.The customer also reported the analyzer generated "vcsn temperature not in operating range limit".A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the system.There were no erroneous test results with this event.There was no death, injury or affect to user or patient treatment.
 
Manufacturer Narrative
(b)(6).On (b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the system and confirmed the thermistor inside the peltier module was defective.The fse replaced the peltier module resolving the issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s kraemer blvd
brea, CA 92821
7149614941
MDR Report Key4004460
MDR Text Key4773910
Report Number1061932-2014-01892
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081930
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Other Device ID NumberSOFTWARE VERSION: 2.0.2.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-