The patient was undergoing a microneurosurgery procedure using the apollo system apollo wand.During the procedure, the apollo wand became clogged and was unable to aspirate clot.The physician covered the tip of the apollo wand to determine if aspiration was occurring and found that it was not aspirating despite the pump working as intended.A new apollo wand was used to complete the procedure and removed a significant volume of clot.Following the procedure, it was noticed that the new apollo wand developed a leak at the plastic y adaptor connecting to the hard black transducer.
|
Result: there is no visible damage to the catheter.Conclusion: the complaint has been evaluated.The complaint indicates that there was no aspiration coming from the apollo catheter/wand.During the evaluation of the returned device, the apollo catheter/wand was flushed with cavicide disinfectant solution.A piece of blood clot was observed in the cavicide disinfectant effluent.The apollo catheter/wand was subsequently tested for aspiration, irrigation, and vibration.The catheter/wand functioned without an issue.It is likely that the flushed clot could have led to this complaint.Once the clot was removed the apollo catheter/wand was functional.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This mdr is associated with mdr 3005168196-2014-00537.
|