SYNTHES TUTTLINGEN IN SITU BENDER-RIGHT; INSTR., BENDING OR CONTOURING Back to Search Results
Catalog Number 388.112
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/23/2014
Event Type  malfunction  
Event Description
It was reported that during surgery utilizing the in-situ benders, the bender tips splayed open.The surgery included 5.5 mm rods in a deformity construct in levels t6-s2.The case resumed with the construct complete, and without delay or complications for the patient.There were no fragments and the patient outcome was good.The damaged tips were first noticed postoperatively, during cleaning of the instruments.This report is for the right side bender this is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional common device name: hxw.Device is an instrument and is not implanted/explanted.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
A review of the device history records has been requested.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.The device history record review could not be performed as the records could not be identified as device is over 9 years old.The visual inspection of the returned device performed as part of the product investigation reported the part was received with the slot opening at the curved end bent open.Material deformation is evident at both openings.One in situ rod bender right (part# 388.112 lot# a7qa28) and one in situ rod bender left (part# 388.113 lot# a7qa28) were returned for the tips reportedly bending open.The benders are included in uss, pangea, and click¿x systems for in situ sagittal plane bending of rods.Alternatively, rod benders 388.91, 388.92, 388.112, 388.113, 388.117, 388.961, 03.620.020, 03.622.060, 03.622.061, and 03.622.062 may be used to achieve contouring of the rods (technique guides: j4437-b, j7243-b, j6418-c, j5462-b, j9009-b, j6718-a).Product drawings 388_112 rev c and 388_113 were reviewed during the investigation.Both instruments have the slot opening, where the instrument is placed around the rod, bent open.The openings have yielded material along the lateral half suggesting that the instrument may not have been fully seated around the rod when used.Multiple areas of material deformation at the opening also indicate that excessive force may have been used repeatedly.Both benders are made from h900 heat treated 17-4ph and is appropriate for the intended use of the instruments as h900 provides maximum hardness.Wear coupled with user technique may have caused the complaint condition however the user technique is unknown and so the root cause is indeterminate.Review of the product drawings found the material, dimensions, tolerances and finishing process of the instruments adequate for their intended use.The complaint may be a result of gradual wear, applying excessive bending force or user technique although it cannot be definitively determined which contributed to the complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN SITU BENDER-RIGHT
Type of Device
INSTR., BENDING OR CONTOURING
Manufacturer (Section D)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen PA 19380 785
GM  19380 7853
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 7853 2
GM   78532
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4013052
MDR Text Key4878567
Report Number2520274-2014-13168
Device Sequence Number1
Product Code HXP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number388.112
Device Lot NumberA7QA28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received09/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-