It was initially reported that the zimmer ats 3000 would deflate during surgery and after disconnecting the cuffs, the ats would inflate.Additional clinical info determined that there was pt involvement associated with the report and there was leakage of blood due to the deflation of the cuff.It was clarified that the amount of blood loss that was observed during the reported event was less than 100cc.There was a delay of around 15 minutes to the surgery and no alternate device was retrieved for use during the procedure.No further clinical info is indicated.
|