The cause of the discrepant falsely elevated aptt results is unknown.Qc at 5:41 was out of range, indicating a potential issue with the reagent vial that had been on board of the instrument for greater than 40hr.Based on internal testing and the complaint investigation, siemens has no evidence for a systematic malfunction of the affected lot.The issue seems correlated with an unknown cause of reduced reagent stability of a single reagent vial while being on board.The issue seems to be a single case at the customer site.Vials tested internally for reagent stability did not shown any deviation from the specifications, and we have no further complaint on the same issue of the affected lot.No further evaluation of the device is required.
|