MEDTRONIC NEUROMODULATION UNKNOWN DEEP BRAIN STIMULATION LEAD; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number NEU_UNKNOWN_LEAD |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/31/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that after the stimloc base plate was screwed into the patient¿s cranium that the slider ¿would not sit and click in the stimloc base plate.¿ the operating physician reportedly ¿tried repositioning the stimloc base plate, however the slider would still not sit in the base plate correctly.¿ a second stimloc was tried at that time and was noted to have ¿worked perfectly.¿ the base plate in question was retained at the time of report, however it was noted the slider had been mistakenly discarded.There was ¿no¿ reported patient injury and the patient was reported to have ¿recovered fully¿ following the incident.A supplemental report will be filed if additional information is received.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id 924256, lot# 082205614a, product type: accessory; product id 924256, lot# 082205614a, product type: accessory.(b)(4), analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.
|
|
Manufacturer Narrative
|
Device evaluation: analysis of the stimloc base found ¿the base worked properly with a known good clip, cap, and lead; however, one of the two screw holes in the base was found to be deformed and molded incorrectly.The hole is smaller near the bottom and larger at the top.The diameter of the hole in both areas is out of specification.It measured 0.075" at the bottom and 0.092" at the top (spec: 0.068"±0.005").¿.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|