• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION UNKNOWN DEEP BRAIN STIMULATION LEAD; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION UNKNOWN DEEP BRAIN STIMULATION LEAD; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number NEU_UNKNOWN_LEAD
Device Problem Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/31/2014
Event Type  malfunction  
Event Description
It was reported that after the stimloc base plate was screwed into the patient¿s cranium that the slider ¿would not sit and click in the stimloc base plate.¿ the operating physician reportedly ¿tried repositioning the stimloc base plate, however the slider would still not sit in the base plate correctly.¿ a second stimloc was tried at that time and was noted to have ¿worked perfectly.¿ the base plate in question was retained at the time of report, however it was noted the slider had been mistakenly discarded.There was ¿no¿ reported patient injury and the patient was reported to have ¿recovered fully¿ following the incident.A supplemental report will be filed if additional information is received.
 
Manufacturer Narrative
Concomitant medical products: product id 924256, lot# 082205614a, product type: accessory; product id 924256, lot# 082205614a, product type: accessory.(b)(4), analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.
 
Manufacturer Narrative
Device evaluation: analysis of the stimloc base found ¿the base worked properly with a known good clip, cap, and lead; however, one of the two screw holes in the base was found to be deformed and molded incorrectly.The hole is smaller near the bottom and larger at the top.The diameter of the hole in both areas is out of specification.It measured 0.075" at the bottom and 0.092" at the top (spec: 0.068"±0.005").¿.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN DEEP BRAIN STIMULATION LEAD
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4024865
MDR Text Key4876727
Report Number3007566237-2014-02334
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNEU_UNKNOWN_LEAD
Device Catalogue NumberNEU_UNKNOWN_LEAD
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00072 YR
-
-