A steris service technician arrived onsite, inspected the washer, and identified that the unit was operating properly.The technician found that the detergent monitoring system, which operates independently from the washer, was not properly detecting the amount of detergent injected during a cycle.The technician identified that the washer cycles completed successfully; the washer can be operated without the detergent monitoring system.The steris service technician purged the tubing which delivers detergent to the washer, recalibrated and primed the system.The unit has been returned to service.The vision single chamber washer/disinfector operator manual states (pp.4-39), "good hospital practice dictates all instruments be inspected for visible debris after processing in the reliance vision single-chamber washer/disinfector or in the reliance vision single- chamber large washer/disinfector.Any instrument with visible debris must be rewashed until clean and free of visible debris prior to terminal processing.Failure to reprocess until all visible debris have been removed may impede the terminal processing.".
|