A surgeon reported during a cataract extraction procedure, a patient injury occurred when the iris was drawn into the phaco tip.The surgeon indicated that it is possible he inadvertently selected the incorrect phaco submode settings.He may have chosen the quad mode settings believing it to be using the peristaltic pump during the sculpt stage which increases the vacuum from 15mmhg to 120mmhg, the surgeon suspected the increase in vacuum may have caused the iris to be drawn despite the phaco tip was within the capsular bag as well as the patient may have had a very floppy iris; however not confirmed.Additional information was received indicating during the procedure, the surgeon had requested an increase in phaco power and instead vacuum was increased pulling iris into the tip.Subsequently, the patient experienced iris damage and as a result the patient has a hole in the iris.Although not confirmed, the surgeon stated the device was incorrectly prepared for use and eye anatomy was the probable cause that contributed to the patient injury.
|