Device evaluated by manufacturer: a visual inspection was performed and the device has a kink at the distal section while it is in a neutral position at 1.5cm from the tip.In addition, the device has broken adhesive at the ring 2 and has blood evidence under the adhesive of the ring 2, and the rings damaged location match with kink location.The steering knob and the tension control knob functioned properly on both lock and unlock positions.Dimensional inspection was performed and the device failed the right and left curve.The device was dissected at the distal section and one steering wire and the center support are broken, at 1.7cm from the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|