BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME® XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004P4790THN40 |
Device Problem
Bent (1059)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/30/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on analysis completed on (b)(4) 2014. it was reported that the device unable to deliver current and catheter bent occurred. the target lesion was located in the cavotricuspid isthmus lesion.A 7-110-2.5-8-10 blazer prime® xp was selected to treat the lesion.However, it was noted that the device was unable to deliver current.The physician checked the catheter under fluoroscopic control and noted that the was catheter bent. the procedure was completed with another of the same device.No patient complications were reported and the patient¿s status was good. however, returned device analysis revealed that the device has a broken adhesive at ring 3.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned for analysis.Visual inspection of the returned device revealed a kink at the distal section while in the neutral position at 2.2cm from the tip.The device has a broken adhesive at ring 3 and blood evidence under the adhesive of the ring 3.The ring damage is in the same location of the kink.The ablation was verified by using the maestro generator 3000 and the device was found within specification.Electrical test was performed, and the device was found within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(6).
|
|
Search Alerts/Recalls
|
|
|