Catalog Number 6605632
Device Problems Intermittent Continuity (1121); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/31/2014
Event Type  malfunction  
Event Description
The customer reported a coulter lh 750 hematology analyzer generated intermittent incomplete computation for the hemoglobin (hgb) parameter results on patient samples.Also, latron control results recovered low out of specification on conductivity.The customer did not provide patient sample results generated during this event.There were no erroneous results reported out of the laboratory and patient treatment was not impacted in connection with this event.
 
Manufacturer Narrative
A beckman coulter (bec) field service engineer (fse) was dispatched to evaluate the instrument.The fse replaced tubing through valves vl4 and vl6.The fse then adjusted a hemoglobin (hgb) lamp, which resolved the intermittent incomplete computation for hgb results.To resolve the low conductivity on latron controls, the fse cleaned the flowcell.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4039078
MDR Text Key4727783
Report Number1061932-2014-02071
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Other Device ID NumberSW VERSION 2D3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/31/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-